|
(Reprinted from Journal
of Electromyography and Kinesiology, volume 6, number
4, pages III-IV, copyright 1996, with permission from Elsevier
Science.)
Authors are advised that the following information must be supplied
in the Methods section of all submitted manuscripts. To avoid delay
or return of manuscripts, the requirements below should be considered
when preparing the manuscript.
Electrodes:
Reported on surface recording of EMG should include:
- electrode material (eg, Ag/AgCl, etc)
- electrode geometry (discs, bars, rectangular, etc)
- size (eg, diameter, radius, width x length)
- use of gel or paste, alcohol applied to cleanse skin, skin
abrasion, shaving of hair, etc
- interelectrode distance
- electrode location, orientation over muscle with respect to
tendons, motor point, and fibers direction
Intramuscular wire electrodes should be described by:
- wire material (eg, stainless steel, etc)
- if single- or multi-strand
- insulation material
- length of exposed tip
- method of insertion (eg, hypodermic needle, etc)
- depth of insertion
- if single or bipolar wire
- location of insertion in the muscle
- interelectrode distance
- type of ground electrode used, location
Needle electrodes and their application should be described
according to standard clinical protocol. The use of nonstandard
needle electrodes should be fully described and include material,
size (gauge), number and size of conductive contact points at the
tip, depth of insertion, and accurate location in the muscle.
Amplification:
Amplifiers should be described by the following:
- if single, differential, double differential, etc
- input impedance
- Common Mode Rejection Ratio (CMRR)
- signal-to-noise ratio
- actual gain range used
Filtering of the raw EMG should be specified by:
- low and/or high pass filters
- filter types (eg, Butterworth, Chebyshev, etc)
- low and/or high pass cut-off frequencies
Since the power density spectra of the EMG contains most of its
power in the frequency range of 5-500 Hz at the extremes, the journal
will not accept reports in which surface EMG was filtered above
10 Hz as a low cut-off and below 350 Hz as the high cut-off: eg,
10-350 Hz is preferred for surface recordings. Filtering
in the band of 10-150 Hz or 50-350 Hz, for example, is not acceptable
as portions of the signal's power above 150 Hz and below 50 Hz
are eliminated. This should be kept in mind when designing a study's
protocol. Exceptions will be made only in rare cases that carry
full scientific justification.
Intramuscular recording should be made with the appropriate
increase of the high frequency cut-off to a minimum 450 Hz. A bandpass
filter of 10-450 Hz is therefore required.
Needle recording should have a bandwidth of 10-1500 Hz.
Rectification: A note should be made if full or half-wave rectification
was carried out.
EMG Processing: There are several methods of EMG processing. Smoothing the
signal with a low pass filter of a given time constant (normally
50-250 ms) is best described as "smoothing with a low-pass
filter of x ms." Alternatively, one can describe it as a "linear
envelope" or "the Mean Absolute Value," while giving
time constant, type, and order of the low-pass filter used. Designating
the EMG subjected to this procedure as the "integrated EMG" (IEMG)
is incorrect (see below).
Another acceptable method is determination
of the "Root Mean
Square" or RMS. Authors should include the time (period) over
which the average RMS was calculated.
Integrated EMG is sometimes reported, but the signal
is actually integrated over time, rather than just smoothed. Such
procedure allows observation of the accumulated EMG activity over
time and should be presented with information as to whether time
or voltage was used to reset the integrator and at what threshold.
Power Density Spectra presentation of the EMG should
include:
- time epoch used for each calculation segment
- type of window used prior to taking the Fast Fourier transform
(FFT) (eg, Hamming, Hanning, Tukey, etc)
- taking the algorithm (eg, FFT)
- number of zero padding applied in the epoch and the resultant
resolution
- equation used to calculate the Median Frequency (MDF), Mean
Frequency (MNF), etc
- the muscle length or fixed joint angle at the time of recording
Other processing techniques, especially novel techniques, are
encouraged if accompanied by full scientific description.
Sampling EMG into the Computer:
Computer processing of the EMG is encouraged if authors observe
these important factors:
1. It is advisable that the raw EMG (eg, after differential amplification
and bandpass filtering) be stored in the computer before further
analysis in case modification of the protocol is required in the
future. In this case, the minimal acceptable sampling rate is at
least twice the highest frequency cut-off of the bandpass filter,
eg, if a bandpass filter of 10-350 Hz was used, the minimal sampling
rate employed to store the signal in the computer should be 700
Hz (350 x 2) and preferably higher to improve accuracy
and resolution. Sampling rates below twice the highest frequency
cut-off will not be accepted.
2. If smoothing with a low-pass filter was performed with hardware
prior to sampling and storing data in the computer, the sampling
rate could be drastically reduced. Rates of 50-100 Hz are sufficient
to introduce smoothed EMG into the computer.
3. It is also advisable that authors consider recording the raw
EMG (prior to bandpass filtering) in the computer. In such cases
a sampling rate of 2500 Hz or above could be used. Yet, to avoid
aliasing of high-frequency noise, bandpass filtering (written in
software) in the range prescribed above should be performed prior
to any further processing of the signal. This approach allows authors
to perform EMG recording with minimal hardware and maximal flexibility.
Yet, it may be at the expense of taxing computer memory space and
speed.
4. Number of bits, model, manufacturer of A/D card used to sample
data into the computer should be given.
Normalization: In investigations where the force/torque was correlated
to the EMG, it is common to normalize the force/torque and its
respective EMG, relative to the values at maximal voluntary contraction
(MVC). Authors should be aware that obtaining true MVC from subjects
requires some preliminary training. Without training, the MVC could
be as much as 20-40% less of that obtained after appropriate training
and lead to incorrect conclusions or interpretation of data. The
journal, therefore, will not accept reports in which subjects were
not properly trained to elicit true MVC.
Normalizing the force/torque with respect to its MVC is commonly
performed with MVC as 100% of the force/torque, and other force
levels are expressed as the appropriate % of MVC. Similarly, the
EMG associated with 100% MVC is designated as 100% and fractions
thereof. Both force/torque and EMG normalization should include
other relevant information such as joint angle(s) and/or muscle
length(s) in isometric contractions, and range of joint angle,
muscle length, velocity of shortening/elongation, and load applied
for non-isometric contractions.
Normalization of data collected from one experimental condition
with respect to other contractile conditions can be performed for
comparative purposes and will be accepted by the journal only if
full description and justification is given in the Methods section.
In sum, the following information should be provided when normalizing
data:
- how subjects were trained to obtain MVC
- joint angle or muscle length
- angles of adjoining joint, eg, for studies on elbow flexion,
the position of the wrist and shoulder joints should be provided
- rate of rise of force
- velocity of shortening/elongation
- changes in muscle length
- ranges of joint angle/muscle length in non-isometric contraction
- load applied in non-isometric contractions
EMG Crosstalk:
Authors should demonstrate that an earnest effort was undertaken
to determine that EMG crosstalk from muscles near the muscle of
interest did not contaminate the recorded signal. Selecting the
appropriate electrode size, interelectrode distance and location
of recordings over the muscle should be carefully planned, especially
when working on areas where many narrow muscles are tightly gathered
(eg, forearm) or when working with superficial/thin muscles (eg,
trapezius). The work of Winter (1994) and Fugelvand et al (1992)
should be consulted if doubts exist. Care also should be employed
when recording surface EMG from areas with subcutaneous adipose
tissue (eg, abdomen, buttocks, chest, etc) as it is known that
adipose tissue enhances crosstalk (Solomonow et al, 1994).
-
Back to Top -
|
|